1SENSIA DR
Report
- Report Number
- 6000144-2011-04336
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S043
- Removal / Correction Number
- Z-2120-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH THRESHOLDS. IT WAS ALSO REPORTED THAT THE PACEMAKER HAD HIGH RV OUTPUTS AND REACHED EARLY ELECTIVE REPLACEMENT INDICATOR (ERI) BECAUSE OF THE RV THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED. THE DEVICE WAS ALSO EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 1SENSIA DR | IMPLANTABLE PULSE GENERATOR | NVZ | MEDTRONIC MED REL, INC. | SEDR01 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD |