FDA Adverse Event Injury Summary report: N

SIGMA 300 D

MDR report key: 2203360 · Received August 11, 2011

Report

Report Number
6000094-2011-01527
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
Z-1510/1519-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS NO PACEMAKER ACTIVITY NOTED ON THE ELECTROCARDIOGRAM, AND THERE WAS NO TELEMETRY. THE DEVICE WAS REMOVED AND ANOTHER WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 300 D IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SD303 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| R (B)(4) COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD