FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2203346 · Received August 11, 2011

Report

Report Number
2649622-2011-12328
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND PRIMARY ANALYSIS REVEALED THAT THE OUTER INSULATION WAS BREACHED (CLAVICLE-RIB CRUSH). THERE WAS BLOOD/BODY FLUID PRESENT ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED). THE OUTER INSULATION WAS BREACHED CUT AND HAD A COSMETIC DEPRESSION. IT WAS ALSO NOTED THAT THE HELIX/LOBE WAS DISTORTED/BENT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND PRIMARY ANALYSIS REVEALED THAT THE OUTER INSULATION WAS BREACHED (CLAVICLE-RIB CRUSH). THERE WAS BLOOD/BODY FLUID PRESENT ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED). THE OUTER INSULATION WAS BREACHED CUT AND HAD A COSMETIC DEPRESSION. IT WAS ALSO NOTED THAT THE HELIX/LOBE WAS DISTORTED/BENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD THRESHOLDS INCREASED SINCE IMPLANT AND "FIBROTIC INGROWTH" WAS NOTED ON THE HELIX. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD