FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2203323
·
Received August 11, 2011
Report
- Report Number
- 2649622-2011-12110
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- June 1, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS ALSO NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND MECHANISM (SLEEVE HEAD) AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE HELIX WOULD NOT EXTEND. A NEW LEAD WAS USED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| 4194 IMPLANTABLE PACING LEAD| 4470 COMPETITOR IMPLANTABLE PACING LEAD |