FDA Adverse Event
Injury
Summary report: N
MAXIMO VR
MDR report key: 2203303
·
Received August 11, 2011
Report
- Report Number
- 6000144-2011-04321
- Event Type
- Injury
- Date Received
- August 11, 2011
- Date of Event
- May 4, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S37
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION, HOWEVER PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS REVEALED A POWER ON RESET (POR) OF THE PARAMETERS. THERE WAS ONE POR FOR WRITE TO LOCKED RAM, ADDR=19A5, DATA=B4, ON (B)(4)-2011 16:31:13 AND ONE PATIENT ALERT FOR POR ON (B)(4)-2011 15:31:13.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE EXPERIENCED AN ELECTRICAL RESET. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO VR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | 7232CX | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention | 6949 IMPLANTABLE TACHY LEAD |