FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2203278 · Received August 11, 2011

Report

Report Number
2649622-2011-11852
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 10, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, THE DEFIBRILLATION COIL WAS DISTORTED. THE INNER TUBING WAS KINKED/BUCKLED. THE HELIX/LOBE WAS DISTORTED/BENT. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND SLEEVE HEAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THE PATIENT'S ANATOMY PRECLUDED THE LEAD FROM BEING IMPLANTED. ACCESS WAS EXTREMELY TIGHT AND THE COIL GOT STUCK AND WOULD NOT ADVANCE. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other