BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL
Report
- Report Number
- 3002773840-2025-00043
- Event Type
- Malfunction
- Date Received
- May 15, 2025
- Date of Event
- April 4, 2025
- Report Date
- May 15, 2025
- Manufacturer
- BIOFIRE DIAGNOSTICS, LLC
- Product Code
- PAM
- UDI-DI
- 00815381020338
- PMA / PMN Number
- K193519
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION: A POTENTIAL FALSE POSITIVE S. MARCESCENS RESULT ON THE BIOFIRE BCID2 PANEL WAS REPORTED AFTER TESTING A BLOOD CULTURE SAMPLE FROM A 77-YEAR-OLD MALE PATIENT. THE CUSTOMER STATED THAT THE PATIENT HAD NO CLINICAL HISTORY. ON (B)(6) 2025, AN INITIAL BLOOD SAMPLE WAS COLLECTED INTO BLOOD CULTURE BOTTLES (BIOMERIEUX BACT/ALERT FA PLUS; LOT: 4102956; EXP: 2025/09/17). ON (B)(6) 2025, THE POSITIVE BLOOD CULTURE WAS TESTED ON THE BIOFIRE BCID2, WHICH RESULTED IN ENTEROBACTERALES, STAPHYLOCOCCUS SPP., AND S. MARCESCENS AS DETECTED. NO GROWTH WAS OBSERVED IN CULTURE. GRAM-POSITIVE COCCI IN PAIRS WERE OBSERVED ON GRAM STAIN. THE CUSTOMER STATED THAT S. MARCESCENS WAS NOT REPORTED TO THE PHYSICIAN. THE PATIENT WAS NOT IMPACTED DUE TO THE BIOFIRE BCID2 PANEL RESULT. NO SERIOUS INJURY OR DEATH WAS REPORTED. CONCLUSION: THE INVESTIGATION DETERMINED THAT THE MOST LIKELY CAUSE OF THE DISCREPANT S. MARCESCENS RESULT WAS THE PRESENCE OF NON-VIABLE ORGANISM/NUCLEIC ACID IN THE BLOOD CULTURE MEDIA BOTTLE. SIMILAR EVENTS WERE RECENTLY OBSERVED IN THE FIELD WITH THE COMBINATION OF THE BIOFIRE BCID2 PANEL AND SPECIFIC LOTS OF BACT/ALERT® CULTURE MEDIA BOTTLES. A FIELD SAFETY CORRECTIVE ACTION (FSCA) WAS ISSUED IN THE U.S. ON JUNE 7, 2024 VIA FSCA FA-TWD-000005. FSCA FA-TWD-000029, AN EXTENSION OF FSCA FA-TWD-000005, WAS ISSUED IN THE U.S. ON JANUARY 22, 2025, FOR THIS ISSUE. WHILE BLOOD CULTURE VIALS ARE AUTOCLAVED AND ROUTINELY QUALITY CONTROLLED FOR STERILITY, NON-VIABLE ORGANISMS OR NUCLEIC ACIDS CAN REMAIN IN THE BLOOD CULTURE MEDIA AFTER THE STERILIZATION PROCESS. THE PRESENCE OF NON-VIABLE ORGANISMS AND NUCLEIC ACID DOES NOT COMPROMISE THE INTENDED USE OF BLOOD CULTURE MEDIA, CULTURING VIABLE MICROORGANISMS; HOWEVER, THE BIOFIRE BCID2 PANEL DOES NOT DISTINGUISH BETWEEN NUCLEIC ACID FROM VIABLE OR NON-VIABLE ORGANISMS. THE "LABORATORY PRECAUTION" AND "LIMITATION" SECTIONS OF THE BIOFIRE BCID2 PANEL INSTRUCTIONS FOR USE (IFU) (WWW.ONLINE-IFU.COM/ITI0048) OUTLINES THE POTENTIAL FOR FALSE POSITIVE DETECTIONS DURING MOLECULAR TESTING WITH STERILE BLOOD CULTURE MEDIA CONTAINING DETECTABLE LEVELS OF NON-VIABLE ORGANISMS AND/OR NUCLEIC ACID. IMPORTANTLY, RESULTS FROM THE BIOFIRE BCID2 PANEL ARE INTENDED TO BE CORRELATED WITH THE CLINICAL HISTORY, EPIDEMIOLOGICAL DATA, AND OTHER DATA AVAILABLE TO THE CLINICIAN EVALUATING THE PATIENT. ALL IDENTIFICATION RESULTS PROVIDED BY THE BIOFIRE BCID2 PANEL ARE INTENDED TO BE INTERPRETED IN CONJUNCTION WITH GRAM STAIN RESULTS. CLINICAL PERFORMANCE: ACCORDING TO TABLE 33. BIOFIRE BCID2 PANEL CLINICAL PERFORMANCE SUMMARY, ENTEROBACTERALES OF THE BIOFIRE BCID2 PANEL IFU, THE PERFORMANCE CLAIM FOR THE S. MARCESCENS ASSAY COMPARED TO STANDARD MANUAL AND AUTOMATED MICROBIOLOGICAL/BIOCHEMICAL IDENTIFICATION METHODS SHOWED AN OVERALL SENSITIVITY OF 100% (95% CI 87.5-100%) AND AN OVERALL SPECIFICITY OF 100% (95% CI 99.7-100%).
SUMMARY: (B)(6) HOSPITAL (B)(6) REPORTED A POTENTIAL FALSE POSITIVE SERRATIA MARCESCENS RESULT ON THE BIOFIRE BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL AFTER TESTING A PATIENT BLOOD CULTURE SAMPLE. THE PATIENT WAS NOT IMPACTED DUE TO THE BIOFIRE BCID2 PANEL RESULT. THE INVESTIGATION DETERMINED THAT THE MOST LIKELY CAUSE OF THE DISCREPANT S. MARCESCENS RESULT WAS THE PRESENCE OF NON-VIABLE ORGANISM/NUCLEIC ACID IN THE BLOOD CULTURE MEDIA BOTTLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1909084 | BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL | BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL | PAM | BIOFIRE DIAGNOSTICS, LLC | RFIT-ASY-0147 | 1808824 | 00815381020338 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Male |