FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2203267 · Received August 11, 2011

Report

Report Number
2649622-2011-12097
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
June 1, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN WAS UNABLE TO PLACE THE LEAD IN THE DESIRED LOCATION. THE LEAD WAS NOT USED AND A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other