FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2203256 · Received August 11, 2011

Report

Report Number
2649622-2011-11672
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS PRESENTED TO THE EMERGENCY ROOM WITH COMPLETE HEART BLOCK AND NO CAPTURE OF THE RIGHT VENTRICULAR LEAD. THE LEAD FRACTURED AND HAD POSITIONAL INTERMITTENT CAPTURE. THE LEAD HAD UNACCEPTABLE THRESHOLDS, HIGH IMPEDANCE, AND OVER SENSING. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization| L| R (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD