FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 2203239 · Received August 11, 2011

Report

Report Number
2649622-2011-11845
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON ALL CONDUCTORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT IMPLANT THE LEAD HAD POSITIONING DIFFICULTIES AND WAS NOT STABLE IN THE DESIRED POSITION. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other