FDA Adverse Event Malfunction Summary report: N

INZII RETRIEVAL SYSTEM

MDR report key: 22032279 · Received May 15, 2025

Report

Report Number
2027111-2025-00535
Event Type
Malfunction
Date Received
May 15, 2025
Date of Event
April 10, 2025
Report Date
November 7, 2025
Manufacturer
APPLIED MEDICAL RESOURCES
Product Code
GCJ
UDI-DI
00607915117382
PMA / PMN Number
K060051
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE EVENT UNIT WAS RETURNED TO APPLIED MEDICAL FOR EVALUATION. VISUAL INSPECTION CONFIRMED THE COMPLAINANT¿S EXPERIENCE OF BEAD DETACHMENT, AS THE BEAD WAS DETACHED FROM THE SPECIMEN BAG. APPLIED MEDICAL HAS REVIEWED THE DETAILS SURROUNDING THE EVENT AND RELATED PRODUCTS AND IS UNABLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT BASED ON THE EVALUATION OF THE RETURNED UNIT. APPLIED MEDICAL HAS PERFORMED A HISTORICAL TREND ANALYSIS AND REVIEW OF PRODUCTION RECORDS AND NO SYSTEMIC PATTERNS WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

THE EVENT UNIT HAS RETURNED TO APPLIED MEDICAL FOR EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 0

PROCEDURE PERFORMED: NA. EVENT DESCRIPTION: COMPLAINT 1 OF 3: 2025-000990, COMPLAINT 2 OF 3: 2025-001076, COMPLAINT 3 OF 3: 2025-001077. DURING A CHOLECISTECTOMY DURING THE MOMENT TO CLOSE THE ENDOBAG WITH THE SPECIMEN INSIDE. THE DOCTOR START TO INSERT THE INZII TO PUT INSIDE THE CHOLECYST SO HE STARTED TO PUSH THE PISTON TO GET THE BAG, HE INSERT THE SPECIMEN, WHEN THEY WERE STARTING TO PULL THE PISTON UP AND TAKE OFF THE GUIDE WIRE THE BAG STUCK IN THE TUBE HOLE. THE DOCTOR TRY SEVERAL TIME TO TAKE DOWN THE BAG AND SO USING STRENGTH THE BAG GET DOWN BUT THE PROBLEM IS THE ROUND GREEN GUIDE WIRE ON THE TOP THAT IS NOT MORE FIXED IN THE WIRE. THE PATIENT IS OK, BUT THE PROBLEM IS THE ROUND GREEN LITTLE BALL THAT ON THE WIRE GUIDE. THIS CAN FALL INTO THE ABDOMEN. THEY FORCED THE BAG FROM THE METAL TUBE AND IN SOME WAYS ( A LOT) THEY FINALLY CLOSED THE BAG AND TAKE IT AWAY. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 16APR25: I HAVE TWO STERILE SAMPLE OPENED FOR ME FROM THE 2 DIFFERENT BOXES TO VERIFY THE PROBLEM, THE THIRD SAMPLE ( THAT THEY USED IN THE OR) THEY GIVE ME ONLY THE PAPER REF BECAUSE THE SAMPLE WAS THREW IT AWAY AND THEY SHOW ME THIS ONE SAMPLE I MEAN (WHAT I CAN SEE IS THE PAPER) BELONG TO ANOTHER BOX. SO TOTALLY THEY HAVE 3 BOXES WITH THE SAME LOT. I HAVE 2 STERILE SAMPLE OPENED IN FRON OF ME FROM 2 OF THESE 3 BOXES TO SHOW ME THE MALFUNCTION OF THEM. THE THIRD BOX I HAVE ONLY THE PAPER WHICH HAVE THE SAME LOT AND CAUSED THE PROBLEM IN THE OPERATING ROOM. SO, ALL THE 3 DEVICES WERE FAULTY, THEY WERE STERILE SAMPLE (NOT ANYMORE WHEN THEY OPEN IN FRONT OF ME). TWO OF THESE WERE OPEN TO ME AS INSPECTION AND ONE WAS DURING A SURGERY/TREATMENT, (THE ONE THAT I PHYSICALLY DON¿T HAVE ONLY HAVE THE PAPER REF) LOT NUMBER IS THE SAME OF THESE 3 CD001:1532774. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 18APR25: YES, THE BEAD DETACHED. THE SNAP RING DIDN'T DETACH. INTERVENTION: NA. PATIENT STATUS: NA (NO PATIENT INVOLVEMENT).

Description of Event or Problem · 0

PROCEDURE PERFORMED: NA EVENT DESCRIPTION: COMPLAINT 1 OF 3: (B)(4), COMPLAINT 2 OF 3: (B)(4), COMPLAINT 3 OF 3: (B)(4). DURING A CHOLECISTECTOMY DURING THE MOMENT TO CLOSE THE ENDOBAG WITH THE SPECIMEN INSIDE. THE DOCTOR START TO INSERT THE INZII TO PUT INSIDE THE CHOLECYST SO HE STARTED TO PUSH THE PISTON TO GET THE BAG, HE INSERT THE SPECIMEN, WHEN THEY WERE STARTING TO PULL THE PISTON UP AND TAKE OFF THE GUIDE WIRE THE BAG STUCK IN THE TUBE HOLE. THE DOCTOR TRY SEVERAL TIME TO TAKE DOWN THE BAG AND SO USING STRENGTH THE BAG GET DOWN BUT THE PROBLEM IS THE ROUND GREEN GUIDE WIRE ON THE TOP THAT IS NOT MORE FIXED IN THE WIRE. THE PATIENT IS OK, BUT THE PROBLEM IS THE ROUND GREEN LITTLE BALL THAT ON THE WIRE GUIDE. THIS CAN FALL INTO THE ABDOMEN. THEY FORCED THE BAG FROM THE METAL TUBE AND IN SOME WAYS ( A LOT) THEY FINALLY CLOSETHE BAG AND TAKE IT AWAY. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 16APR25: I HAVE TWO STERILE SAMPLE OPENED FOR ME FROM THE 2 DIFFERENT BOXES TO VERIFY THE PROBLEM, THE THIRD SAMPLE ( THAT THEY USED IN THE OR) THEY GIVE ME ONLY THE PAPER REF BECAUSE THE SAMPLE WAS THREW IT AWAY AND THEY SHOW ME THIS ONE SAMPLE I MEAN ( WHAT I CAN SEE IS THE PAPER) BELONG TO ANOTHER BOX. SO TOTALLY THEY HAVE 3 BOXES WITH THE SAME LOT. I HAVE 2 STERILE SAMPLE OPENED IN FRONT OF ME FROM 2 OF THESE 3 BOXES TO SHOW ME THE MALFUNCTION OF THEM. THE THIRD BOX I HAVE ONLY THE PAPER WHICH HAVE THE SAME LOT AND CAUSED THE PROBLEM IN THE OPERATING ROOM. SO, ALL THE 3 DEVICES WERE FAULTY, THEY WERE STERILE SAMPLE (NOT ANYMORE WHEN THEY OPEN IN FRONT OF ME) TWO OF THESE WERE OPEN TO ME AS INSPECTION AND ONE WAS DURING A SURGERY/TREATMENT, (THE ONE THAT I PHYSICALLY DON¿T HAVE ONLY HAVE THE PAPER REF) LOT NUMBER IS THE SAME OF THESE 3 CD001 :1532774. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 18APR25: YES, THE BEAD DETACHED. THE SNAP RING DIDN'T DETACH. INTERVENTION: NA PATIENT STATUS: NA (NO PATIENT INVOLVEMENT).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1215933 INZII RETRIEVAL SYSTEM LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ APPLIED MEDICAL RESOURCES CD001 1532774 00607915117382

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown