FDA Adverse Event
Malfunction
Summary report: N
UNICEL ¿ DXI 800 ACCESS ® IMMUNOASSAY SYSTEM
MDR report key: 2203211
·
Received August 11, 2011
Report
- Report Number
- 2122870-2011-02898
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 12, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4)
Additional Manufacturer Narrative · 1
ON (B)(6), 2011, A BEC FIELD SERVICE ENGINEER (FSE) DISCOVERED LEAKING WASH BUFFER TRANSFER PERI-PUMP TUBING IN THE FLUIDICS DRAWER. THE FSE REPLACED ALL OF THE PERI-PUMP TUBING. HARDWARE IS THE ROOT CAUSE FOR THIS EVENT. BEC INTERNAL IDENTIFICATION # IS (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) IN REGARDS TO A FLUID LEAK INTO THE WASTE DRAWER UNDER THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. THE FLUID LEAK FROM THE PERI-PUMPS IS MOST LIKELY WASH BUFFER. CUSTOMER TECHNICAL SUPPORT (CTS) RECOMMENDED THE USE OF PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) TO CLEAN UP THE SPILL. NO EXPOSURE OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL ¿ DXI 800 ACCESS ® IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | DXI 800 | N/A |