FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 2203209 · Received August 11, 2011

Report

Report Number
2649622-2011-11840
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S10
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION AND THE DEVICE MET EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET AND WAS AT ELECTIVE REPLACEMENT INDICATOR. THE DEVICE WAS EXPLANTED AND REPLACED. THE RIGHT VENTRICULAR LEAD HAD HIGH THRESHOLD AND LOW IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED. DURING THE IMPLANT, AN ADDITIONAL LEAD WAS ATTEMPTED, BUT NOT IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4023 ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| R 4523 IMPLANTABLE PACING LEAD