FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2203175 · Received August 11, 2011

Report

Report Number
2649622-2011-11832
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. 1 - PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(6) 2011. 2 - VENTRICULAR NST=200 MS AVERAGE V-CYCLE ON (B)(6) 2011. VENTRICULAR SHORT INTERVAL COUNT V-SIC=228 COUNTS, IN 0.05 DAY, ON (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT WAS TRIGGERED AND THERE WAS NOISE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB