FDA Adverse Event Injury Summary report: N

SELECT SECURE

MDR report key: 2203170 · Received August 11, 2011

Report

Report Number
2649622-2011-11830
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P030036
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION HAD A COSMETIC CUT, THERE WAS APPARENT EXPLANT DAMAGE, AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION HAD A COSMETIC CUT AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

INFECTION WAS REPORTED. THE ORGANISM IDENTIFIED IS (B)(6). THE LEADS WERE REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SELECT SECURE IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 3830 ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| R D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD