FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 2203146 · Received August 11, 2011

Report

Report Number
6000144-2011-04309
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS REVEALED NORMAL BATTERY DEPLETION AND THE DEVICE MEETS 80% OF EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD PREMATURE BATTERY DEPLETION. THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 4193 IMPLANTABLE PACING LEAD| 4068 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| 4024 IMPLANTABLE PACING LEAD