FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR AT

MDR report key: 2203097 · Received August 11, 2011

Report

Report Number
6000094-2011-01510
Event Type
Injury
Date Received
August 11, 2011
Date of Event
January 28, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. THE DEPLETED BATTERY WAS CAUSED BY HIGH CURRENT LEAKAGE IN THE C45 TANTALUM CAPACITOR.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S.

Description of Event or Problem · 1

IT WAS REPORTED THAT REPLACEMENT RECOMMENDED TIME INDICATOR TRIGGERED ON THE DEVICE AND A PATIENT ALERT OCCURRED. SINCE THE DEVICE HAD ONLY BEEN IMPLANTED FOR APPROXIMATELY FOUR MONTHS, PREMATURE BATTERY DEPLETION SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT REPLACEMENT RECOMMENDED TIME INDICATOR TRIGGERED ON THE DEVICE AND A PATIENT ALERT OCCURRED. SINCE THE DEVICE HAD ONLY BEEN IMPLANTED FOR APPROXIMATELY FOUR MONTHS, PREMATURE BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC S.A. C174AWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R