FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2203052 · Received August 11, 2011

Report

Report Number
1423500-2011-10598
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
July 22, 2011
Report Date
August 8, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE ASSIGNABLE CAUSE OF THE IIPV IDENTIFIED VIA DEVICE LOG REVIEW WAS DETERMINED TO BE INSUFFICIENT DRAIN, USE ERROR, TIDAL ULTRAFILTRATION (UF) REMOVAL SET TOO LOW. DEVICE MET SPECIFICATIONS RELATIVE TO IIPV IN THE LOGS.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2011 DURING CYCLE 5. THE PATIENT'S ULTRAFILTRATION (UF) READING WAS 1527 ML, INDICATING THE HOME PATIENT (HP) DRAINED 1527 ML MORE THAN THEIR PROGRAMMED FILL VOLUME OF 2200 ML. THIS INFORMATION GAVE A TOTAL DRAIN VOLUME OF 3727 ML, WHICH MEETS IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 53 YR