HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-10598
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- July 22, 2011
- Report Date
- August 8, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE ASSIGNABLE CAUSE OF THE IIPV IDENTIFIED VIA DEVICE LOG REVIEW WAS DETERMINED TO BE INSUFFICIENT DRAIN, USE ERROR, TIDAL ULTRAFILTRATION (UF) REMOVAL SET TOO LOW. DEVICE MET SPECIFICATIONS RELATIVE TO IIPV IN THE LOGS.
(B)(4). THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2011 DURING CYCLE 5. THE PATIENT'S ULTRAFILTRATION (UF) READING WAS 1527 ML, INDICATING THE HOME PATIENT (HP) DRAINED 1527 ML MORE THAN THEIR PROGRAMMED FILL VOLUME OF 2200 ML. THIS INFORMATION GAVE A TOTAL DRAIN VOLUME OF 3727 ML, WHICH MEETS IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |