ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2011-05807
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Report Date
- July 13, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THERE IS NO COMPLAINT AGAINST THE PUMP (WHICH LISTED IN THE INITIAL REPORT). THE COMPLAINT SHOULD BE AGAINST THE CARTRIDGE LISTED IN THE INITIAL REPORT.
THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT # B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE. SERIAL #(B)(4) REFERS TO THE INSULIN PUMP INVOLVED IN THE COMPLAINT. CATALOG # = 100-124-01. RECALL 2531779-02/25/11-001-R.
ON (B)(6) 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ALLEGING THAT A CARTRIDGE HAD POSSIBLY LEAKED. THE PATIENT ALSO REPORTED BLOOD GLUCOSE (BG) VALUES OF "424 MG/DL" AND "327 MG/DL" WHICH WERE OBTAINED DURING THE TIME OF CONCERN. THE PATIENT REPORTEDLY TOOK A MANUAL INJECTION OF INSULIN AT AN UNSPECIFIED TIME. THE PATIENT DENIED HAVING DEVELOPED SYMPTOMS OF NAUSEA, VOMITING, CHEST PAIN, AND SHORTNESS OF BREATH. DURING A CARTRIDGE CHANGE, THE PATIENT CLAIMED THAT SHE NOTICED INSULIN IN THE CARTRIDGE CHAMBER. HE ALSO CLAIMED THAT HE HAD ENCOUNTERED THE SAME LEAKING CARTRIDGE ISSUE 8 WEEKS EARLIER. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED HE HAD A LEAKING CARTRIDGE ISSUE. THERE IS NO EVIDENCE, HOWEVER, THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY SYMPTOMS, TREATMENT, OR BLOOD GLUCOSE VALUES SUGGESTIVE OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART | B201581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |