FDA Adverse Event Injury Summary report: N

CAPSURE Z NOVUS

MDR report key: 2203004 · Received August 11, 2011

Report

Report Number
2649622-2011-11597
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD HAD HIGH IMPEDANCE. IT WAS ALSO REPORTED THERE WERE SENSING AND PACING FAILURE AS CONFIRMED BY ELECTROGRAM. AN APPARENT LEAD FRACTURE WAS FOUND BY X-RAY. USE OF THE ATRIAL LEAD WAS STOPPED AND THE PACEMAKER SETTING WAS CHANGED TO VVI MODE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE Z NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5554 ASKU

Patients

Seq Age Sex Outcome Treatment
1 88 YR Required Intervention (B)(4) IMPLANTABLE PULSE GENERATOR| 5054 IMPLANTABLE PACING LEAD