FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2202937 · Received August 11, 2011

Report

Report Number
2649622-2011-11572
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 20, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN INITIAL IMPLANT, THE HELIX OF THE LEAD WOULD NOT FIXATE AFTER SEVERAL POSITIONING ATTEMPTS. THE HELIX WOULD RETRACT, BUT TO EXTEND THE HELIX TOOK OVER FIFTY ROTATIONS. WHEN THE HELIX DID FIXATE, THE PACING AND SENSING THRESHOLDS WERE POOR. THE LEAD WAS NOT USED AND A NEW LEAD IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Other