ADAPTA DR
Report
- Report Number
- 6000094-2011-01405
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S043
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS OPENED AND THE DEVICE WAS NOTED TO HAVE HIGHER THAN NOMINAL CURRENT. THE HIGH CURRENT DRAIN WAS SHOWN TO BE CAUSED BY A FAULTY XT122 TANTALUM CAPACITOR (C1 BATTERY FILTER CAPACITOR).
IT WAS REPORTED THAT PREMATURE BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADAPTA DR | IMPLANTABLE PULSE GENERATOR | NVZ | MEDTRONIC S.A. | ADDRS1 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R | 4592 IMPLANTABLE PACING LEAD| 4092 IMPLANTABLE PACING LEAD| 4092 IMPLANTABLE PACING LEAD| 4592 IMPLANTABLE PACING LEAD |