FDA Adverse Event
Injury
Summary report: N
ENRHYTHM DR
MDR report key: 2202927
·
Received August 11, 2011
Report
- Report Number
- 6000144-2011-04044
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. POWER ON RESET - POWER ON RESET PARAMETERS. CRITICAL RAM PARITY ERROR LOGGED ON (B)(6) 2011 IN (B)(6). POR SEVERITY IS CONSIDERED LOW AS DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER RESET.
Description of Event or Problem · 1
IT WAS REPORTED THE PACEMAKER ELECTRICALLY RESET REQUIRING REPROGRAMMING. THE PACEMAKER AND LEADS ARE STILL IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Required Intervention | (B)(4) IMPLANTABLE PACING LEAD| 4024 IMPLANTABLE PACING LEAD |