FDA Adverse Event Injury Summary report: N

SKYTRON LLC OR TABLES

MDR report key: 22028920 · Received May 15, 2025

Report

Report Number
MW5170430
Event Type
Injury
Date Received
May 15, 2025
Date of Event
May 5, 2025
Report Date
May 12, 2025
Manufacturer
SKYTRON, LLC.
Product Code
GDC
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

OPERATING ROOM 5 REPORTED FIRE FROM THE SKYTON OPERATION TABLE MODEL 3602 SAW A SPARK FROM THE BED, AND SHE THINKS HER BED WAS ON FIRE. BIOMED WAS ON SITE TO ANALYZE THE ISSUE AND IT WAS FOUND THAT THE TABLE WAS WET AROUND THE POWER CODE AREA. THERE WAS NO HARM TO THE PATIENT AND STAFF ON THIS REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88707 SKYTRON LLC OR TABLES TABLE, OPERATING-ROOM, ELECTRICAL GDC SKYTRON, LLC. 3602

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other| R