FDA Adverse Event Malfunction Summary report: N

S5 CONSOLE

MDR report key: 22025614 · Received May 15, 2025

Report

Report Number
9611109-2025-90120
Event Type
Malfunction
Date Received
May 15, 2025
Date of Event
April 15, 2025
Report Date
May 20, 2026
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DTQ
UDI-DI
04033817902676
PMA / PMN Number
K210130
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A.1.-A.5. THERE WAS NO PATIENT INVOLVEMENT. H11: LIVANOVA DEUTSCHLAND MANUFACTURES THE S5 CONSOLE PRO VERSION. THE INCIDENT OCCURRED IN INDIA. IN ORDER TO SOLVE THE ISSUE, MOTOR CONTROL BOARD (HMS) AND MOTOR POWER AMPLIFIER BOARD (HMF) NEED TO BE REPLACED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

H11: A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO REPLACE MOTOR CONTROL BOARD (HMS) AND MOTOR POWER AMPLIFIER BOARD (HMF), HOWEVER THE ISSUE REMAINED. DURING FURTHER TROUBLESHOOTING, IT WAS FOUND THAT THE ISSUE WAS WITH THE MOTOR SHAFT. THEREFORE, THE TWO NEW BOARDS WERE REMOVED, REINSTALLED THE OLD ONES AND SENT THE INVOLVED AFFECTED PUMP TO LIVANOVA DEUTSCHLAND FOR A FURTHER INVESTIGATION AND REPAIR. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A S5 CONSOLE PRO VERSION GAVE A MOTOR CONTROL FAILURE ERROR MESSAGE (CODE 442) DURING PRIMING, AFTER INSTALLATION OF THE MACHINE. RESTARTING THE MACHINE AND RECONNECTING ALL CABLES DID NOT RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
614566 S5 CONSOLE CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS DTQ LIVANOVA DEUTSCHLAND 48-40-00Z 04033817902676

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown