Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2202539
·
Received August 5, 2011
Report
- Report Number
- 2032227-2011-01964
- Event Type
- Injury
- Date Received
- August 5, 2011
- Date of Event
- July 22, 2011
- Report Date
- July 24, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 425 MG/DL. IT WAS STATED THAT THE CANNULA WAS BENT WHEN THE INFUSION SET WAS REMOVED. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP PASSED THE PRIME AND HIGH PRESSURE TESTS. THE CALLER REQUESTED AN INSERTION DEVICE FOR THE CUSTOMER DUE TO BENT CANNULAS WHEN INSERTING MANUALLY. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-722LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR | Hospitalization |