FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2202497 · Received August 5, 2011

Report

Report Number
3004209178-2011-82442
Event Type
Injury
Date Received
August 5, 2011
Date of Event
July 17, 2011
Report Date
July 17, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A BUTTON ERROR ALARM AS A RESULT OF CORRODED KEYPAD TRACES. ALSO THE KEYPAD OVERLAY HAD WEAK ADHESIVE BOND AT SEVERAL LOCATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER WAS NOT ABLE TO CLEAR THE ALARM. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. DURING THE CALL, THE CUSTOMER MENTIONED THAT HER RECERTIFIED INSULIN PUMP DID MALFUNCTION AND SHE HAD TO GO TO THE HOSPITAL. ATTEMPTED TO CALL THE CUSTOMER TO GET MORE INFO REGARDING HER ADMISSION, AND HER HUSBAND STATED THAT SHE WENT TO THE HOSPITAL OVER A YEAR AGO DUE TO HER INSULIN PUMP MALFUNCTIONING, BUT THE CUSTOMER DID NOT RECEIVE A MEDICAL TREATMENT. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization