FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 22024656 · Received May 15, 2025

Report

Report Number
2955842-2025-20485
Event Type
Malfunction
Date Received
May 15, 2025
Date of Event
April 18, 2025
Report Date
April 21, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
10886874115661
PMA / PMN Number
K173337
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE VESSEL SEALER EXTEND INSTRUMENT WAS ANALYZED AND THE COMPLAINT WAS NOT CONFIRMED BY FAILURE ANALYSIS. THE INSTRUMENT WAS PLACED AND DRIVEN ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION, ENGAGEMENT, CUT AND SEALED AS EXPECTED. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS OPENED AND CLOSED PROPERLY. THERE WERE NO FAILURES REPORTED IN THE LOGS. NO PRODUCT ISSUE WAS IDENTIFIED. DEVICE HISTORY RECORD (DHR) REVIEW FOR THE DEVICE(S) INVOLVED WITH THE REPORTED EVENT HAS BEEN COMPLETED. NO NON-CONFORMANCES WERE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SIGMOID COLECTOMY SURGICAL PROCEDURE, THE SURGEON REPORTED THE VESSEL SEALER EXTEND INSTRUMENT SEEMED TO BE MALFUNCTIONING AND NOT MIRRORING HIS MOVEMENTS. IT SEEMED AS IF THE CABLES WERE NOT WORKING PROPERLY. THE INSTRUMENT WOULD NOT ARTICULATE PROPERLY. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER TO OBTAIN ADDITIONAL INFORMATION, HOWEVER, NO FURTHER DETAILS ARE AVAILABLE REGARDING THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347454 ENDOWRIST VESSEL SEALER EXTEND NAY INTUITIVE SURGICAL, INC 480422-03 L18241205 0328 10886874115661

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES