FDA Adverse Event Malfunction Summary report: N

PATROL PUMP

MDR report key: 2202252 · Received August 5, 2011

Report

Report Number
1527460-2011-00053
Event Type
Malfunction
Date Received
August 5, 2011
Report Date
July 6, 2011
Manufacturer
ABBOTT NUTRITION
Product Code
LZH
PMA / PMN Number
K943733
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE TESTING AND INVESTIGATION RESULTS CONFIRMED AN UNDER-DELIVERY WITH A STUTTERING MOTOR. THE POLE PIECES OF THE MOTOR WERE FOUND TO BE MISALIGNED. THE FRONT PUMP CASE WAS BROKEN. ABBOTT NUTRITION STANDARD PROCEDURE INCLUDES ATTEMPTS TO OBTAIN ALL REQUIRED INFO FROM THE SOURCE.

Description of Event or Problem · 1

THE COMPLAINANT REPORTED AN UNDER-DELIVERY. THE INTENDED FEEDING RATE WAS 45 ML PER HR AND THE PUMP DELIVERED 200 ML OVER 12 HRS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PATROL PUMP LZH, PUMP, INFUSION, ENTERAL LZH ABBOTT NUTRITION 52036

Patients

Seq Age Sex Outcome Treatment
1