FDA Adverse Event
Malfunction
Summary report: N
PATROL PUMP
MDR report key: 2202252
·
Received August 5, 2011
Report
- Report Number
- 1527460-2011-00053
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Report Date
- July 6, 2011
- Manufacturer
- ABBOTT NUTRITION
- Product Code
- LZH
- PMA / PMN Number
- K943733
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE TESTING AND INVESTIGATION RESULTS CONFIRMED AN UNDER-DELIVERY WITH A STUTTERING MOTOR. THE POLE PIECES OF THE MOTOR WERE FOUND TO BE MISALIGNED. THE FRONT PUMP CASE WAS BROKEN. ABBOTT NUTRITION STANDARD PROCEDURE INCLUDES ATTEMPTS TO OBTAIN ALL REQUIRED INFO FROM THE SOURCE.
Description of Event or Problem · 1
THE COMPLAINANT REPORTED AN UNDER-DELIVERY. THE INTENDED FEEDING RATE WAS 45 ML PER HR AND THE PUMP DELIVERED 200 ML OVER 12 HRS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PATROL PUMP | LZH, PUMP, INFUSION, ENTERAL | LZH | ABBOTT NUTRITION | 52036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |