FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2202244 · Received August 11, 2011

Report

Report Number
2649622-2011-11308
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS FRACTURED. IT WAS NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED, THE OUTER INSULATION HAD A COSMETIC CUT, THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A CARELINK TRANSMISSION INDICATED THE RIGHT VENTRICULAR (RV) LEAD HAD A LEAD WARNING AND IMPEDANCES GREATER THAN 9,999 OHMS AND NO CAPTURE. THE PATIENT WAS SEEN AT THE CLINIC AND A LEAD REVISION WAS SCHEDULED. WHEN THE LEAD WAS EVALUATED WITH THE ANALYZER DURING THE RV LEAD REVISION THE R-WAVES WERE INCONSISTENT, IMPEDANCE WAS GREATER THAN 4000 OHMS AND THERE WAS NO CAPTURE AT 5.0 VOLTS. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD