FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2202193 · Received August 5, 2011

Report

Report Number
1720753-2011-20736
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 21, 2011
Report Date
August 5, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP REPLACED THE HIGH VOLTAGE TANK, GENERAL DRIVER, FILAMENT DRIVER, ALIGNED THE HIGH VOLTAGE SUPPLY REGULATOR, FILAMENT CALIBRATION AND VERIFIED THE DOSAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED LOW KILOVOLTS AND THE JOYSTICK DIDN'T WORK. THIS HAPPENED DURING A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1