FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2202172 · Received August 5, 2011

Report

Report Number
1720753-2011-20755
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
August 1, 2011
Report Date
August 5, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE KEY WAS EXTRACTED FROM THE SLOT. THE KEY WAS REPLACED WITH A CUSTOMER KEY. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE KEY BROKE OFF IN THE SYSTEM. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1