FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 2202159 · Received August 5, 2011

Report

Report Number
1720753-2011-20766
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 28, 2011
Report Date
August 5, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP CONDUCTED AN ONSITE INVESTIGATION. THE COLLIMATOR WAS CALIBRATED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A COLLIMATOR POTENTIOMETER ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1