FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2202128 · Received August 11, 2011

Report

Report Number
2182208-2011-01200
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 19, 2011
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE REPORTED EVENT. TELEMETRY WAS INTERMITTENT. THE PRINTED CIRCUIT BOARD AND ANTENNA WERE FOUND TO BE OUT OF SPECIFICATION. THE CASE HANDLE WAS ALSO NOTED TO BE BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT TELEMETRY WAS POOR OR INTERMITTENT WITH DEVICES. PRESSURE ON THE PROGRAMMER RF HEAD CONNECTOR WOULD SOMETIMES HELP ESTABLISH TELEMETRY. THE PROGRAMMER HEAD WAS REPLACED, WITH NO CHANGE. BOOT UP WAS ALSO SLOW. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other