FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2202128
·
Received August 11, 2011
Report
- Report Number
- 2182208-2011-01200
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- May 19, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE REPORTED EVENT. TELEMETRY WAS INTERMITTENT. THE PRINTED CIRCUIT BOARD AND ANTENNA WERE FOUND TO BE OUT OF SPECIFICATION. THE CASE HANDLE WAS ALSO NOTED TO BE BROKEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT TELEMETRY WAS POOR OR INTERMITTENT WITH DEVICES. PRESSURE ON THE PROGRAMMER RF HEAD CONNECTOR WOULD SOMETIMES HELP ESTABLISH TELEMETRY. THE PROGRAMMER HEAD WAS REPLACED, WITH NO CHANGE. BOOT UP WAS ALSO SLOW. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |