FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2202097 · Received August 11, 2011

Report

Report Number
2649622-2011-11274
Event Type
Injury
Date Received
August 11, 2011
Date of Event
May 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVEHEAD, THE LEAD WAS STRETCHED, AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVEHEAD, AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT EXPERIENCED SEVERE CONTRACTIONS/CONVULSIONS POST-IMPLANT. THE DEVICE AND TWO LEADS WERE REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| R