CAPSUREFIX NOVUS

MDR report key: 2202039 · Received August 11, 2011

Report

Report Number
2649622-2011-11265
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 3, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THERE WAS BLOOD IN/ON THE HELIX LOBE MECHANISM, AND THE INNER INSULATION WAS KINKED AND BUCKLED. THE LEAD APPEARED TO HAVE BEEN STRETCHED AND DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THERE WERE MULTIPLE ATTEMPTS TO POSITION THE RIGHT VENTRICULAR (RV) LEAD, EACH RESULTING IN HIGH THRESHOLDS WITH SENSING DIFFICULTY. ADDITIONALLY, THE STYLET GOT STUCK IN THE LEAD. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Other