FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2202017 · Received August 11, 2011

Report

Report Number
2649622-2011-11261
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. OVERSENSING, VENTRICULAR (NON-SUSTAINED TACHYCARDIA) NST LESS THAN OR EQUAL TO 210 MS AVERAGE V-CYCLE ON (B)(6) 2011 IN THE TIMEFRAME. VENTRICULAR FIBRILLATION EQUAL 200 MS AVERAGE V-CYCLE ON (B)(6) 2011. INTERFERENCE/NOISE, VENTRICULAR SHORT INTERVAL COUNT (V-SIC) EQUALS 1267.3 COUNTS AVG/DAY, IN 9.90 DAYS, BETWEEN (B)(6) 2011. HIGH RESISTANCE/IMPEDANCE AND PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2011. DAILY PACE IMPEDANCE TREND DATA SHOWS AN ABRUPT SPIKE INCREASE FOR RV PACE EQUAL TO 472 TO INF (UNFILTERED DATA) OHMS PEAK BETWEEN (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED MULTIPLE INAPPROPRIATE SHOCKS DUE TO AN APPARENT RIGHT VENTRICULAR (RV) LEAD FRACTURE. AN EVALUATION OF THE RV LEAD SHOWED HIGH LEAD IMPEDANCE AND THAT A PATIENT ALERT WAS TRIGGERED. THE LEAD HAD OVERSENSING AND NOISE WITH ARM MOVEMENT. THE PATIENT DESCRIBES PULLING THEMSELVES UP WITH THEIR LEFT ARM TO GET INTO A VEHICLE AND SOON AFTER THE SHOCKS STARTED TO OCCUR. THE LEAD WAS CAPPED AND REPLACED BY A NEW LEAD. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| L| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD