SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2011-11260
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- May 4, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- Z-0474-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD/BODY FLUID IN THE DISTAL CONDUCTOR (NOT OBSTRUCTED). THERE WAS ALSO BLOOD IN/ON THE HELIX MECHANISM AND SLEEVE HEAD. THE TIP SEAL WAS OBSERVED OUT OF POSITION.
IT WAS REPORTED THAT A RIGHT VENTRICULAR LEAD IMPLANTATION WAS ATTEMPTED, BUT NOT SUCCESSFUL BECAUSE THE HELIX WOULD NOT EXTEND. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Other |