FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2202016 · Received August 11, 2011

Report

Report Number
2649622-2011-11260
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 4, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
Z-0474-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD/BODY FLUID IN THE DISTAL CONDUCTOR (NOT OBSTRUCTED). THERE WAS ALSO BLOOD IN/ON THE HELIX MECHANISM AND SLEEVE HEAD. THE TIP SEAL WAS OBSERVED OUT OF POSITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A RIGHT VENTRICULAR LEAD IMPLANTATION WAS ATTEMPTED, BUT NOT SUCCESSFUL BECAUSE THE HELIX WOULD NOT EXTEND. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935 ASKU

Patients

Seq Age Sex Outcome Treatment
1 32 YR Other