FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX

MDR report key: 2201988 · Received August 11, 2011

Report

Report Number
2649622-2011-11252
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 9, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): DISTAL CONDUCTOR DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE DISTAL CONDUCTOR FRACTURED DUE TO OVERSTRESS, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED DAMAGE AT IMPLANT; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, WHILE THE LEAD WAS BEING REPOSITIONED, THE PHYSICIAN RETRACTED THE HELIX BUT IT WOULD NOT LATER EXTEND. IT WAS ALSO NOTED THAT THE STYLET SEEMED TO NEED AN EXCESSIVE AMOUNT OF FORCE TO REMOVE IT FROM THE LEAD. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5568 ASKU

Patients

Seq Age Sex Outcome Treatment
1 93 YR Other (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD