FDA Adverse Event
Injury
Summary report: N
VASO VIEW HEMOPRO
MDR report key: 2201976
·
Received August 5, 2011
Report
- Report Number
- 2242352-2011-01070
- Event Type
- Injury
- Date Received
- August 5, 2011
- Date of Event
- July 18, 2011
- Report Date
- July 19, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. WE WILL CONTINUE PURSUING THE DEVICE BEING RETURNED BY THE CUSTOMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. ITEMS MARKED "NI" ARE UNK TO US AT THIS TIME. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO PLASTIC SIDE OF THE C-RING BROKE OFF INTO THE PT'S LEG; THIS IS THE SCOPE WASH TUBING. THE USER HAD TO MAKE AN ADD'L SURGICAL CUTDOWN TO REMOVE THE PIECE OF THE DEVICE. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. MAQUET CARDIOVASCULAR ANTICIPATES RETURN OF THE PRODUCT IN QUESTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASO VIEW HEMOPRO | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Required Intervention |