ENRHYTHM DR
Report
- Report Number
- 6000144-2011-03852
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION.
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR NORMAL BATTERY DEPLETION, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Hospitalization| R | 4092 IMPLANTABLE PACING LEAD| 4592 IMPLANTABLE PACING LEAD |