FDA Adverse Event Injury Summary report: N

PROTECTA XT VR

MDR report key: 2201933 · Received August 11, 2011

Report

Report Number
6000094-2011-01343
Event Type
Injury
Date Received
August 11, 2011
Date of Event
May 10, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(6) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN THE ELECTROPHYSIOLOGIST WAS DOING THE IMPEDANCE TEST, TWAVE OVERSENSING (TWOS) WAS OBSERVED. AFTER REPROGRAMMING, TWOS DISAPPEARED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROTECTA XT VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. D354VRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 6935 IMPLANTABLE TACHY LEAD