SPRINT
Report
- Report Number
- 2649622-2011-11209
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S13
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED A LEAD INTEGRITY ALERT TRIGGERED - 1 PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(4)-2011 09:33:55. OVERSENSING - THERE WERE 2 VENTRICULAR NST=210 MS AVERAGE V-CYCLE ON (B)(4)-2011 17:59:51 AND (B)(4)-2011 02:27:30. INTERFERENCE/NOISE - VENTRICULAR SHORT INTERVAL COUNT V-SIC=2.1 COUNTS AVG/DAY, IN 138.42 DAYS, BETWEEN (B)(4)-2010 AND (B)(4)-2011 11:08:29.
IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT TRIGGERED DUE TO SENSING INTEGRITY COUNTS AND NON-SUSTAINED TACHYCARDIA EVENTS. OVERSENSING WAS REPRODUCED WITH ISOMETRIC PATIENT MOVEMENTS. A RIGHT VENTRICULAR LEAD FRACTURE WAS NOTED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6945 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |