FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 2201845 · Received August 11, 2011

Report

Report Number
2649622-2011-11209
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED A LEAD INTEGRITY ALERT TRIGGERED - 1 PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(4)-2011 09:33:55. OVERSENSING - THERE WERE 2 VENTRICULAR NST=210 MS AVERAGE V-CYCLE ON (B)(4)-2011 17:59:51 AND (B)(4)-2011 02:27:30. INTERFERENCE/NOISE - VENTRICULAR SHORT INTERVAL COUNT V-SIC=2.1 COUNTS AVG/DAY, IN 138.42 DAYS, BETWEEN (B)(4)-2010 AND (B)(4)-2011 11:08:29.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT TRIGGERED DUE TO SENSING INTEGRITY COUNTS AND NON-SUSTAINED TACHYCARDIA EVENTS. OVERSENSING WAS REPRODUCED WITH ISOMETRIC PATIENT MOVEMENTS. A RIGHT VENTRICULAR LEAD FRACTURE WAS NOTED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6945 ASKU

Patients

Seq Age Sex Outcome Treatment
1 54 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB