FDA Adverse Event Injury Summary report: N

ENTRUST VR

MDR report key: 2201828 · Received August 11, 2011

Report

Report Number
6000144-2011-03820
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
Z-1172-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE DEVICE WAS RETURNED AND ANALYZED AND WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. ADDITIONALLY, THERE WAS A RANDOM ACCESS MEMORY (RAM) CHIP MEMORY ERROR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD ELECTRICAL RESET WITH ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154VRC ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD