FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 2201748 · Received August 11, 2011

Report

Report Number
2649622-2011-11185
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S003
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD FRACTURED AND THAT THERE WAS HIGH IMPEDANCE, OVERSENSING, AND INCREASED THRESHOLDS. CAPTURE COULD NOT BE MAINTAINED, AND THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION. IT WAS NOTED ON FLUOROSCOPY THAT THE LEAD PIN MIGHT BE PULLED BACK INTO THE CONNECTOR BLOCK. HOWEVER, AFTER RESEATING THE PIN THE HIGH IMPEDANCE, NO CAPTURE, AND STIMULATION CONTINUED. THE LEAD WAS CAPPED AND REPLACED, AND THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4193 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| O| R 6949 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD