FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2201560 · Received August 11, 2011

Report

Report Number
2649622-2011-11135
Event Type
Injury
Date Received
August 11, 2011
Date of Event
April 30, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND; HOWEVER, THERE WAS APPARENT EXPLANT DAMAGE. THE OUTER INSULATION HAD A BREACHED CUT AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND; HOWEVER, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD DISLODGED. THE LEAD WAS EXPLANTED AND A DIFFERENT LEAD WAS ATTEMPTED; HOWEVER THE LEAD REQUIRED OVER TWENTY TURNS TO EXTEND THE HELIX. THE LEAD WAS NOT IMPLANTED AND A DIFFERENT LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| R