FDA Adverse Event
Malfunction
Summary report: N
TW POWER SUPPLY, US
MDR report key: 2201464
·
Received August 4, 2011
Report
- Report Number
- 2242352-2011-01151
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 2, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- HQO
- PMA / PMN Number
- K043155
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
METHOD - THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. WE WILL CONTINUE PURSUING THE DEVICE BEING RETURNED BY THE CUSTOMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. (B)(4).
Description of Event or Problem · 1
THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE TW POWER SUPPLY, US WAS INTERMITTENT. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS. MAQUET CARDIOVASCULAR ANTICIPATES RETURN OF THE PRODUCT IN QUESTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TW POWER SUPPLY, US | ENDOSCOPIC VESSEL HARVESTING SYSTEM | HQO | MAQUET CARDIOVASCULAR, LLC | VH-3010 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |