FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 2201423 · Received August 11, 2011

Report

Report Number
2649622-2011-11111
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0068-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE DISTAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR FRACTURED. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS FRACTURED, THE OUTER TUBING OVERLAY WAS MELTED, OUTER TUBING OVERLAY WITH COSMETIC ENVIRONMENTAL STRESS CRACK, OUTER TUBING OVERLAY WAS BREACHED CUT, THE HELIX WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THE LEAD WAS FLEXED, WITHIN FIVE CENTIMETERS OF THE ANCHORING SLEEVE. THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

THE LEAD WAS REMOVED DUE TO AN INFECTION. THE LEAD WAS RETURNED, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6931 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R