FDA Adverse Event Malfunction Summary report: N

TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM

MDR report key: 2201298 · Received August 3, 2011

Report

Report Number
1124841-2011-00340
Event Type
Malfunction
Date Received
August 3, 2011
Date of Event
July 14, 2011
Report Date
July 14, 2011
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
Product Code
GCJ
PMA / PMN Number
K083194
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TERUMO RECEIVED THE ACTUAL DEVICE; UPON EVAL, IT WAS CONFIRMED THAT THE INTERNAL LENS WAS BROKEN, RESULTING IN A BLURRY VISUAL FIELD. VISUAL INSPECTION NOTED MINOR SCRATCHES ON THE ROD AND EYEPIECE OF THE ENDOSCOPE. THIS TYPE OF DAMAGE IS OFTEN ATTRIBUTED TO HANDLING ISSUES. THE DEVICE IFU STATES THAT THE ENDOSCOPE MUST BE HANDLED WITH EXTREME CARE TO PREVENT DAMAGE TO THE DEVICE. (B)(4). ALL AVAILABLE INFO HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING AND F/U.

Description of Event or Problem · 1

THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR SYSTEMS THAT DURING ENDOSCOPIC VEIN HARVESTING, THE ENDOSCOPE WAS BLURRY. THE PRODUCT WAS CHANGED OUT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM ENDOSCOPE GCJ TERUMO CARDIOVASCULAR SYSTEMS CORPORATION MCENDO550 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK