FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2201274 · Received August 3, 2011

Report

Report Number
1720753-2011-20670
Event Type
Malfunction
Date Received
August 3, 2011
Date of Event
July 21, 2011
Report Date
August 3, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP CONDUCTED AN ONSITE INVESTIGATION. TWO CIRCUIT BOARDS WERE REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT DISPLAY AN IMAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1